The objective of this work has been to study the origin of surface failures in UHMWPE tibial bearings of total knee replacements. Earlier examination of hundreds of retrieved prostheses, along with analysis of the properties of UHMWPE material from retrieved bearings, had shown that a large number of the bearings suffered from subsurface oxidation. This oxidation was related to the gamma irradiation used to sterilize the tibial bearings. Mechanical properties of the polymer were significantly deteriorated in the oxidized region, with the most severe reduction of strength and ductility occurring about 1 mm beneath the contact surface. In this work the contact stress distribution in the bearings was analyzed, and tribotesting of the bearing materials was carried out under simulated service conditions. Fatigue cracks and delamination developed in bearings tested in a knee simulator and in rolling/sliding test specimens, and the damage was similar to that found in retrieved tibial bearings. The fatigue cracks invariably initiated in the embrittled oxidized layer, and the depth of that layer determined the depth at which the cracks began. The stress analysis showed that the maximum shear stress and von Mises equivalent stress reached high levels in the subsurface oxidized zone where the delamination and contact fatigue failures initiated. [S0742-4787(00)04401-5]

1.
Cameron, H. U., 1994, Preface, Bone Implant Interface, Robert Hurley, ed., Mosby Year Book, Inc.
2.
ISO, 1996, “Test Methods for Laboratory Simulation of In-Vivo Wear of Total Knee Prostheses,” ISO/NP 14243-1.
3.
Seireg
,
A.
, and
Arvikar
,
R. J.
,
1975
, “
The Prediction of Muscular Load Sharing and Joint Forces in the Lower Extremities During Walking
,”
J. Biomech.
,
8
, pp.
89
102
.
4.
McNamara
,
J. L.
,
Collier
,
J. P.
,
Mayor
,
M. B.
, and
Jensen
,
R. E.
,
1994
, “
A Comparison of Contact Pressures in Tibial and Patellar Total Knee Components Before and After Service In-Vivo
,”
Orthop. Relat. Res.
,
299
, pp.
104
113
.
5.
Bartel
,
D. L.
,
Bicknell
,
V. L.
, and
Wright
,
T. M.
,
1986
, “
The Effect of Conformity, Thickness and Material on Stresses in Ultra-High Molecular Weight Components for Total Joint Replacement
,”
J. Bone Jt. Surg., Am.
,
68A
, pp.
1041
1051
.
6.
Landy
,
M. M.
, and
Walker
,
P. S.
,
1988
, “
Wear of Ultra High Molecular Weight Polyethylene Components of 90 Retrieved Knee Prostheses
,”
J. Arthroplasty Suppl.
,
3
, pp.
S73–S85
S73–S85
.
7.
Sutula
,
L. C.
,
Collier
,
J. P.
,
Saum
,
K. A.
,
Currier
,
B. H.
,
Currier
,
J. H.
,
Sanford
,
W. M.
,
Mayor
,
M. B.
,
Wooding
,
R. E.
,
Sperling
,
D. K.
,
Williams
,
I. R.
,
Kasprzakm
,
D. J.
, and
Surprenant
,
V. A.
,
1995
, “
Impact of Gamma Sterilization on Clinical Performance of Polyethylene in the Hip
,”
Orthop. Relat. Res.
,
319
, pp.
28
40
.
8.
Collier
,
J. P.
,
Sperling
,
D. K.
,
Currier
,
J. H.
,
Sutula
,
L. C.
,
Saum
,
K. A.
, and
Mayor
,
M. B.
,
1996
, “
Impact of Gamma Sterilization on Clinical Performance of Polyethylene in the Knee
,”
J. Arthroplasty
,
11
, pp.
377
389
.
9.
Williams
,
I. R.
,
Mayor
,
M. B.
, and
Collier
,
J. P.
,
1998
, “
The Impact of Sterilization Method on Wear in Knee Arthoplasty
,”
Clin. Orthop. Relat. Res.
,
356
, pp.
170
180
.
10.
Blunn
,
G. W.
,
Joshi
,
A. B.
,
Minns
,
R. J.
,
Lidgren
,
L.
,
Lilley
,
P.
,
Ryd
,
L.
,
Engelbrecht
,
E.
, and
Walker
,
P. S.
,
1997
, “
Wear in Retrieved Condylar Knee Arthroplastics
,”
J. Arthroplasty
,
12
, pp.
281
290
.
11.
ASTM, 1996, “Standard Test Method for Tensile Properties of Plastics,” Standard D638-96.
12.
Kurtz, S. M., Rimnac, C. M., Li, S., and Bartel, D. L., 1994, “A Bilinear Material Model for UHMWPE Components,” Proc. 40th Ann. Meeting of Orthopaedic Res. Soc., p. 289.
13.
Kurtz, S. M., Rimnac, C. M., and Bartel, D. L., 1996, “The Degradation Rate of UHMWPE in Total Joint Replacements,” Proc. 42nd Ann. Meeting of Orthopaedic Res. Soc., p. 492.
14.
Walker
,
P. S.
,
Blunn
,
G. W.
, and
Lilley
,
P. A.
,
1996
, “
Wear Testing of Materials and Surfaces for Total Knee Replacements
,”
J. Biomed. Mater. Res.
,
33
, pp.
159
175
.
15.
Currier
,
J. H.
,
Duda
,
J. L.
,
Sperling
,
D. K.
,
Collier
,
J. P.
,
Currier
,
B. H.
, and
Kennedy
,
F. E.
,
1998
, “
In-Vitro Simulation of Contact Fatigue Damage Found in UHMWPE Components of Knee Prostheses
,”
IMechE J. Eng. Med.
,
212H
, pp.
293
302
.
16.
Kennedy
,
F. E.
,
Lee
,
P. H.
,
Feehan
,
J. P.
, and
Collier
,
J. P.
,
1991
, “
Failure of UHMWPE Polymer Surfaces of Knee Prostheses: A Rolling Sliding Contact
,”
Advances in Engineering Tribology
, Y-W Chung and H. S. Cheng, eds., STLE
SP-31
, pp.
100
109
.
17.
Saum, K. A., 1994, “Oxidation Versus Depth and Time for Polyethylene Gamma Sterilized in Air,”Proc. 40th Ann. Meeting of Orthopaedic Res. Soc., p. 237.
18.
Sun
,
D. C.
,
Stark
,
C.
, and
Dumbleton
,
J. H.
,
1996
, “
Development of an Accelerated Aging Method for Evaluation of Long-Term Irradiation Effects on Ultrahigh Molecular Weight Polyethylent Implants,” Irradiation of Polymers, eds., R. L. Clough and S. W. Shalaby
,
American Chemical Soc.
,
620
, pp.
340
349
.
19.
Blunn
,
G. W.
,
Walker
,
P. S.
,
Joshi
,
A.
, and
Hardinge
,
K.
,
1991
, “
The Dominance of Cyclic Sliding in Producing Wear in Total Knee Prostheses
,”
Clin. Orthop.
,
273
, pp.
253
263
.
20.
Sathasivam
,
S.
, and
Walker
,
P. S.
,
1998
, “
Computer Model to Predict Subsurface Damage in Tibial Inserts of Total Knees
,”
J. Orthop. Res.
,
16
, pp.
564
571
.
21.
Estupinan
,
J. A.
,
Bartel
,
D. L.
, and
Wright
,
T. M.
,
1998
, “
Residual Stresses in Ultra-High Molecular Weight Polyethylene Loaded Cyclically by a Rigid Moving Indenter in Nonconforming Geometries
,”
J. Orthop. Res.
,
16
, pp.
80
88
.
22.
Reeves
,
E. A.
,
Barton
,
D. C.
,
Fitzpatrick
,
D. P.
, and
Fisher
,
J.
,
1998
, “
A Two-Dimensional Model of Cyclic Strain Accumulation in Ultra-High Molecular Weight Polyethylene Knee Replacements
,”
IMechE J. Eng. Med.
,
212H
, pp.
189
198
.
23.
Plumet, S., 1998, “Modelisation d’un Milieu Multicouche 3D Sous Sollicitations de Contact: Application aux Prosthe`ses de Genou Ste´rilise´es,” Doctoral thesis, INSA, Lyon France.
24.
Plumet
,
S.
, and
Dubourg
,
M. C.
,
1998
, “
A 3D Model for a Multilayered Body Loaded Normally and Tangentially Against a Rigid Body: Application to Specific Coatings
,”
ASME J. Tribol.
,
120
, pp.
668
676
.
25.
Kalkar, J. J., 1990, Three Dimensional Elastic Bodies in Rolling Contact, Kluwer Academic Publishers, Dordrecht.
26.
Johnson, K. L., 1985, Contact Mechanics, Cambridge University Press, Cambridge, UK.
You do not currently have access to this content.