Abstract
A tube-load model is used to reconstruct aortic pressure waveform from peripheral pressure waveform. Yet, the reconstructed aortic pressure waveform is greatly affected by load impedance used. In this work, a vibrating-string model for closed-loop wave transmission and reflection between the aorta and periphery is developed to examine the roles of all the parameters involved in aortic pressure waveform. The arterial pulsatile wave theory gives rise to the standard one-dimensional wave equation for a vibrating string. A vibrating-string model based on radial displacement of the arterial wall is developed to relate aortic pressure waveform to peripheral pressure waveform, relate load impedance to input impedance, and derive theoretical expressions for associated clinical indices. The vibrating-string model is extended to incorporate blood velocity and is further connected to the left ventricle (LV) to study the role of the LV in aortic pressure waveform. The difference between the vibrating-string model and the tube-load model is also examined. Load impedance is identified as an indispensable independent parameter for reconstruction of aortic pressure waveform with accuracy, and its physiologically realistic harmonic dependence can only be obtained from the measured input impedance. The derived expressions for clinical indices interpret some clinical findings and underscore the role of harmonics in clinical indices. Some misconceptions in the tube-load model are revealed, including load impedance and characteristic impedance. This work clarifies the role of harmonics-dependence of load impedance and harmonics of aortic pressure waveform in determining clinical indices.
1 Introduction
As compared with peripheral pressure waveform, aortic pressure waveform is more relevant to the left ventricle (LV) function but is difficult and costly to measure [1,2]. Therefore, one-dimensional (1D) models for pulsatile wave propagation in the arterial tree have been developed to reconstruct aortic pressure waveform from measured peripheral pressure waveform [1–7]. Generally speaking, there are two types of 1D models: distributed 1D model and tube-load model [2]. The distributed 1D model includes the arterial tree network (e.g., a large number of arterial segments and bifurcations), and thus involves great computational complexity and has not been extensively used to reconstruct aortic pressure waveform [2]. Built upon a transmission-line analogy, the tube-load model offers great computational simplicity and has been extensively employed in reconstructing aortic pressure waveform [3–7]. In the simplest tube-load model, a uniform, lossless tube represents the arterial section between the aorta and periphery (i.e., the measured artery site), and a load at periphery represents the impedance from the rest arteries and termination [2].
Aortic pressure waveform is characterized by three clinical indices: return time, reflection magnitude, and augmentation index (AI) [2]. As compared with arterial stiffness (i.e., pulse wave velocity (PWV)), these clinical indices have shown their independent values, although they are all thought to be indicative of arterial stiffness [2,5]. In essence, AI is a composite indicator of return time and reflection magnitude. Reconstruction of aortic pressure waveform with accuracy is important for accurate estimates of these clinical indices. To improve accuracy in reconstructed aortic pressure waveform, various complexities (e.g., tapered arterial geometry, transmission loss, and different load impedance) have been incorporated into the tube-load model [2–7]. To date, the majority of the studies on the tube-load model have focused on exploring different load impedance to improve accuracy of reconstructed aortic pressure waveform, because load impedance has significant influence on the shape of reconstructed aortic pressure waveform [1–3,5]. In contrast, input impedance at the aorta has been neglected in the studies on the tube-load model [2–7]. While load impedance separates the forward waves from the reflected waves in peripheral pressure waveform, input impedance separates the forward waves from the reflected waves in aortic pressure waveform. As such, there might be some relationship between the two. It is worth noting that there are measured data for input impedance in the literature [8], but no studies have reported on the measured data for load impedance, to the best knowledge of the author.
Westerhof et al. [1] examined the influence of load type in the tube-load model on input impedance. When a pure resistor is used as the load, the tube-load model fails to interpret the measured relation of return time to PWV in clinical studies and matches the measured input impedance in clinical studies. When a three-element Windkessel model is used as the load, the tube-load model can interpret the measured relation of the return time to PWV and provide reasonable input impedance. However, a theoretical relation of load impedance to input impedance in the tube-load model was not provided. Neither a pure resistor nor a three-element Windkessel model as the load is capable of reconstructing aortic pressure waveform with accuracy [4,5]. Moreover, both the distributed 1D model [1] and the tube-load model [2–7] fall short of clarifying physical implications underlying the influence of the load type on input impedance and aortic pressure waveform.
Aortic pressure waveform is a collection of harmonics (nω, with n and ω as the n-th harmonic and the fundamental frequency, respectively,) of the heartbeat and the sum of the forward and reflected pressure waves [1,5]. Lower-level harmonics are sufficient to capture aortic pressure amplitude but are insufficient to accurately capture aortic pressure waveform [9]. Using machine-learning algorithms, a recent study has found that two sets of values for the three-element Windkessel model as the load for lower-level harmonics and higher-level harmonics, respectively, improve accuracy in reconstructed aortic pressure waveform [5]. This manifests harmonic-dependence of load impedance and insufficiency of one three-element windkessel model for representing this harmonic-dependence. Yet, physical implications of harmonic-dependence of load impedance is unclear, and physiologically realistic harmonic-dependence of load impedance is still unfathomable.
To reconstruct aortic pressure waveform with accuracy, it is essential to clarify physical implications of load type and harmonic-dependence of load impedance and examine the relation of load impedance to input impedance. In this work, built upon the standard 1D wave equation [10], a vibrating-string model based on radial displacement of the arterial wall is developed to clarify the entangled relations of the three parameters: load impedance (ZL), input impedance (Z0), and harmonics (nω), in closed-loop wave transmission and reflection between the aorta and periphery, in which input impedance at the aorta is related to load impedance at periphery. Clarification of the entangled relations reveals the role of the three parameters in aortic pressure waveform and its clinical indices. This vibrating-string model is further extended to incorporate blood velocity. Given the identified role of harmonics of aortic pressure waveform in its clinical indices, the vibrating-string model is connected to the LV to study the LV-artery interaction for exploring the relation of aortic pressure waveform to the driving force on the LV and possible affecting factors of harmonics of aortic pressure waveform. A comparison of the vibrating-string model with the tube-load model is also conducted to identify some misconceptions in the tube-load model for reconstruction of aortic pressure waveform.
The rest of the paper is organized as follows: In Sec. 2, the vibrating-string analogy for arterial pulsatile wave propagation is presented to derive equivalent linear density and equivalent tension from arterial properties and geometries. In Sec. 3, a vibrating-string model based on radial wall displacement is developed to relate aortic pressure waveform to peripheral pressure waveform and relate load impedance to aortic input impedance. Afterwards, return time and reflection magnitude at the aorta is related to both input impedance and load impedance. By relating the theoretical input impedance to the measured input impedance in clinical studies, physiologically realistic harmonic-dependence of load impedance is revealed. The vibrating-string model is extended to incorporate blood velocity. By connecting the vibrating-string model to the LV, a lumped-element mechanical model for the LV-artery interaction is created to relate the driving force on the LV to aortic pressure waveform and explore possible affecting factors of harmonics of aortic pressure waveform. In Sec. 4, the difference between the tube-load model and the vibrating-string model is examined. In Sec. 5, independent parameters and dependent parameters are identified in closed-loop wave transmission and reflection between the aorta and periphery, revealing that load impedance is an indispensable independent parameter for reconstruction of aortic pressure waveform with accuracy, and its physiologically realistic harmonic-dependence can only be obtained from the measured input impedance; and input impedance is a dependent parameter and is determined by load impedance and arterial properties and geometries. The derived expressions for clinical indices and their calculated values, based on the measured harmonics of aortic pressure and blood flow waveforms in the literature [8], interpret some clinical findings and underscore the role of harmonics of aortic pressure waveform in determining its clinical indices. With the vibrating-string model, physical implications of load type and harmonic-dependence of load impedance are made evident. Some misconceptions in the tube-load model become conspicuous. Section 6 summarizes key insights on reconstruction of aortic pressure waveform and its associated physical implications.
2 Vibrating-String Analogy for Pulsatile Wave Propagation in an Artery
In this section, the theory of pulsatile wave propagation in an artery is briefly presented to obtain the standard 1D wave equation for pulsatile parameters in an artery [10]. By comparing this equation with the wave equation for a vibrating string, equivalent tension and equivalent linear density for an artery are related to arterial properties and geometries.
2.1 Pulsatile Wave Propagation in an Artery.
As shown in Fig. 1, the arterial wall is assumed to be a thin-walled tube and has two geometrical parameters: thickness h and radius a at diastolic blood pressure (DBP). The arterial wall undergoes radial displacement η(x, t) during a pulse cycle. Blood is assumed to be incompressible, Newtonian fluid, and undergo laminar flow. There are three pulsatile parameters in blood flow: radial blood velocity w(r, x, t) and axial blood velocity u(r, x, t), and pulsatile pressure Δp(x, t). Each pulsatile parameter is a collection of harmonics (nω) of the heartbeat. When the wavelength of pulsatile pressure is well below the arterial radius [11], pulsatile pressure does not vary in the radial direction.
where with being the Womersley number and [11]; kn = nω/cn is the nth wave number; and cn is the nth wave velocity. Note that Δp0 and B are two constant unknowns.
Note that c0 is the same as PWV in clinical studies and is independent of harmonics. Given the harmonics-dependence of F10, the wave velocity varies with harmonics.
where u(x,t) is the blood velocity averaged across the lumen. While wave reflection augments pulsatile pressure and radial wall displacement, it reduces blood velocity and wall shear stress.
2.2 Standard One-Dimensional Wave Equation for Pulsatile Wave Propagation in an Artery.
While T is harmonics-dependent, ρL is harmonics-independent. Note that 1−F10 is harmonics-dependent at small arteries, such as the carotid artery and the radial artery, and 1−F10 ≅1 is harmonics-independent at the aorta, due to its large size.
3 A Vibrating-String Model for Closed-Loop Wave Transmission and Reflection Between the Aorta and Periphery
According to Eqs. (3a) and (8), radial wall displacement waveform is identical to pressure waveform. Given that radial wall displacement carries evident physical implication to wave transmission and reflection in the context of the 1D wave propagation, radial wall displacement is analyzed here, instead of pulsatile pressure. By modeling the aorta-periphery section as a uniform vibrating-string, aortic radial wall displacement waveform is related to peripheral radial wall displacement waveform, and input impedance is related to load impedance. Additionally, return time and reflection magnitude at the aorta is related to both input impedance and load impedance. The role of harmonics of radial wall displacement on return time and reflection magnitude is identified. By relating the theoretical input impedance to the measured input impedance in clinical studies, physiological realistic harmonics-dependence of load impedance is revealed. The analysis on radial wall displacement is extended to blood velocity as equivalent transverse displacement, and the input power at the aorta for radial wall displacement and blood velocity is derived. By connecting the vibrating-string model to the LV, the LV-artery interaction is analyzed to relate driving force on the LV to aortic pressure waveform.
3.1 Relation of Aortic Radial Wall Displacement Waveform to Peripheral Radial Wall Displacement Waveform.
Except arterial geometries and properties (i.e., a, h, L, E, μ, and ρb) and c0, η(x, t) and other parameters (i.e., load impedance, input impedance, reflection coefficient, wave velocity, and F10) involved in this section are all harmonics-dependent.
3.2 Relation of Input Impedance to Load Impedance.
As will be seen later on, since ϕn < 0 for the first three harmonics (n = 1, 2, 3), the driving force is ahead of the radial wall velocity for these harmonics.
Accordingly, the nth reflection coefficient can be obtained from the nth normalized input impedance and wave velocity (i.e., kn and γn), while the nth load impedance can be obtained from the nth normalized input impedance and wave velocity, as well as the nth characteristic impedance.
3.3 Return Time and Reflection Magnitude and Reconstruction of Aortic Radial Wall Displacement Waveform.
Accordingly, return time varies with harmonics and is affected by the phase of reflection coefficient at periphery, and can be obtained from the input impedance.
As such, the input impedance and the harmonics of the forward wave need to be treated as two independent parameters for determining aortic radial wall displacement waveform. Alternatively, the normalized input impedance and aortic radial wall displacement waveform can be treated as two independent parameters to determine the harmonics of the forward waves.
Evidently, harmonics of the forward waveform affects reflection amplitude for the whole waveform and AI, which is defined as the ratio of augmented pressure to pulsatile pressure (see Fig. 7).
The known values of kn and γn and L are insufficient to reconstruct aortic radial wall displacement waveform from its peripheral counterpart. Either normalized input impedance or reflection coefficient at periphery is needed. Harmonics-dependence of normalized input impedance (or harmonics-dependence of reflection coefficient at periphery) translates to 20 unknown values in Eq. (29) if only the first ten harmonics are used.
3.4 Relation of the Theoretical Input Impedance to the Measured Input Impedance in Clinical Studies.
3.5 Blood Velocity and Input Power at the Aorta.
Accordingly, pulsatile pressure, radial wall displacement, and blood velocity are all analogous to transverse displacement in a vibrating string and propagate along the axial direction with the same wave velocity and characteristic impedance. While radial wall displacement is associated with elastic energy transmission through the arterial wall, blood velocity is associated with kinetic energy transmission and mass transport through blood flow.
As transverse displacement, blood velocity is the sum of the forward waves and the reflected waves in Eq. (34a). The mechanical impedance and input impedance for blood velocity are the same as those for radial wall displacement, as expressed in Eqs. (14) and (32). Comparison of Eqs. (13b) and (34b) shows that the driving force for blood velocity is 2cn/a times larger than that for radial wall displacement.
The total nth input power at the aorta is the sum of Pn-η and Pn-u. Given Pn-u ≫Pn-η, blood flow is the dominant carrier of power transmission. Although the amplitudes of the harmonics at the aorta drop greatly with higher n-level [1,8,10], harmonics-dependent input power might indicate that the contribution of higher n-level harmonics to the input power is non-negligible. Table 1 summarizes the vibrating-string models for radial wall displacement and blood velocity and their comparison with the tube-load model, as described later on.
Vibrating-string model | Tube-load model | ||
---|---|---|---|
Arterial wall | Blood flow | Arterial wall+ blood flow | |
Governing equation | |||
Transverse displacement | η | u | |
Velocity | ∂η/∂t | ∂u/∂t | |
Driving force | −Τ∂η/∂x | −Τ∂u/∂x | |
characteristic impedance | |||
Input impedance | |||
Input power |
Vibrating-string model | Tube-load model | ||
---|---|---|---|
Arterial wall | Blood flow | Arterial wall+ blood flow | |
Governing equation | |||
Transverse displacement | η | u | |
Velocity | ∂η/∂t | ∂u/∂t | |
Driving force | −Τ∂η/∂x | −Τ∂u/∂x | |
characteristic impedance | |||
Input impedance | |||
Input power |
3.6 Left Ventricle-Artery Interaction for Relating Driving Force on the Left Ventricle to Aortic Pressure Waveform.
The input impedance Z0n can be obtained from the measured input impedance. Note that radial wall displacement waveform is identical to pulsatile pressure waveform, and Fη(t) and Fu(t) carry the same waveform, with Fu(t) ≫Fη(t).
4 Comparison of the Tube-Load Model With the Vibrating-String Model
The difference between the uniform tube-load model and the uniform vibrating-string model is examined here.
4.1 Difference in Governing Equations.
According to Eq. (40), pulsatile wave propagation in an artery is analogous to a transmission line: Q and Δp are equivalent to current and voltage, respectively. Due to the assumption on η(x, t) = 0, wave velocity c0 does not vary with harmonics, and damping coefficient R/L is a constant.
When η(x, t) ≠ 0, wall shear stress is harmonics-dependent and is not linearly proportional to Q. Therefore, damping coefficient is not a constant in Eq. (42a). As shown in Eq. (42b), the contribution of wall shear stress to the wave propagation can be included into the x-derivative of pulsatile pressure, leading to a complex, harmonics-dependent wave velocity.
4.2 Difference in Characteristic Impedance, Load Impedance, and Input Impedance.
Table 2 summarizes the difference between the tube-load model with a three-element Windkessel model as the load and the vibrating-string model. Since a pure resistor load is totally physiologically unrealistic (see Sec. 5), it is not included in the table.
Tube-load model | Vibrating-string model | |
---|---|---|
Driving force | Δpn | |
Velocity | Qn | |
Arterial compliance | ||
Arterial inductance | ||
Wave velocity | ||
Damping coefficient | ||
Characteristic impedance | ||
Load impedance | ||
Reflection coefficient | ||
Input impedance |
Tube-load model | Vibrating-string model | |
---|---|---|
Driving force | Δpn | |
Velocity | Qn | |
Arterial compliance | ||
Arterial inductance | ||
Wave velocity | ||
Damping coefficient | ||
Characteristic impedance | ||
Load impedance | ||
Reflection coefficient | ||
Input impedance |
4.3 Difference in Input Power at the Aorta and Driving Force on the Left Ventricle.
5 Discussion
In the tube-load model [2–7], the distributed 1D model [1], and related clinical studies on the CV system [1], Eq. (38) is the theoretical basis for defining the parameters that relate wave transmission and reflection to arterial properties and geometries. Pulsatile pressure and blood flow rate are defined as driving force and velocity, respectively, as shown in Table 2. Then, the ratio of pulsatile pressure versus blood flow rate (without wave reflection) becomes characteristic impedance. Yet, since Eq. (38) does not comply with the standard 1D wave equation, these definitions are not consistent with their counterparts in the context of 1D wave propagation, and their physical implications to wave transmission and reflection are then obscured. Built on the standard 1D wave equation, the vibrating-string model for the arterial tree allows defining of all the parameters consistent with their well-established physical implications in the acoustical field [12], which can be directly adopted for studying wave transmission and reflection in the aorta-periphery section.
Other than the difference in definitions of the parameters and their physical implications between the tube-load model and the vibrating-string model, the vibrating-string model is further utilized to derive theoretical relation of load impedance to input impedance, which reveals that 1) input impedance is a dependent parameter and is determined by load impedance and arterial properties and geometries and 2) the use of a three-element Windkessel model as the load stipulates harmonics-dependence of load impedance, which is inconsistent with the measured input impedance. Moreover, theoretical expressions for return time and reflection magnitude at each harmonic are derived in terms of input impedance and can be calculated based on the measured input impedance.
5.1 Who Are Independent Parameters and Dependent Parameters in Closed-Loop Wave Transmission and Reflection Between the Aorta to Periphery?
Since the aorta is at the start of the arterial tree, it is commonly assumed that aortic pressure waveform is the input and peripheral pressure waveform is the output in the tube-load and distributed 1D models [1–7]. As shown in Eq. (25), since aortic reflected pressure waveform is affected by all the downstream parameters, it should not be included in the input. As shown in Fig. 4, by following the wave transmission path, the input and the output and the role of all the parameters become clear. The input is aortic forward waveform ηf(0,t,nω), which is not affected by the downstream parameters (Note that the LV is excluded here.), and is a collection of harmonics (nω) of the heartbeat. This forward waveform transmits toward periphery, with transmission characteristics (i.e., Zc, c, and L) determined by arterial properties and geometries. At periphery, the forward waveform ηf(L,t,nω) is reflected with load impedance ZL. The reflected waveform ηb(L,t,nω) transmits toward the aorta with the same transmission characteristics, and becomes the reflected waveform at the aorta ηb(0,t,nω), which is the output. Given its influence on small arteries, the fluid-loading term F10 should not be neglected in the aorta-periphery section, and thus harmonics-dependence of wave velocity and characteristic impedance needs to be included in reconstruction of aortic pressure waveform.
Although input impedance Z0 can be calculated from pulsatile parameters at the aorta using Eq. (31), it is determined by transmission characteristics and load impedance, as expressed in Eq. (20). Thus, Z0 is not an independent parameter. Then, four parameters: Zc, c, L, and ZL, are needed to reconstruct aortic pressure waveform from peripheral pressure waveform with accurate representation of the harmonics of aortic pressure waveform. In clinical studies, aortic pressure waveform and blood flow waveform are measured to separate aortic forward waveform from aortic reflected waveform and also calculate Z0. Certainly, measurement at the aorta involves high cost and great technical complexity. In contrast, measurement of peripheral pressure waveform can be easily achieved using a tactile sensor or a photoplethysmographic (PPG) sensor [2–7]. However, measurement of blood flow waveform at peripheral arteries becomes challenging, due to their small size. As such, it is difficult, if not impossible, to measure ZL. As analyzed here, the measured Z0 can be utilized to obtain ZL.
Following the wave transmission path leads to closed-loop wave transmission and reflection between the aorta and periphery: wave transmission starts and ends at the aorta, and wave reflection at the aorta is a combination of round-trip wave transmission and wave reflection at periphery. This closed-loop wave transmission and reflection reveals that the input and the output are both at the aorta and why input impedance is not an independent parameter but depends on load impedance and transmission characteristics. This is important, in the sense that the reflected waveform at the aorta is not an independent parameter, but depends on the forward waveform and load impedance, and transmission characteristics. Then, three clinical indices at the aorta are indicative of load impedance and transmission characteristics, as well as harmonics of the forward waveform. As compared with following the whole waveform along the arterial tree, closed-loop wave transmission and reflection serves better for revealing the entangled relations of input impedance, load impedance, and harmonics in the aorta-periphery section. As seen in Sec. 3, relating input impedance to load impedance allows the calculation of return time and reflection magnitude at the aorta using input impedance, since the latter manifests round-trip wave transmission and wave reflection at periphery.
5.2 Characteristic Impedance and Load Impedance Versus Tapered Arterial Geometry.
In the context of 1D wave propagation, characteristic impedance has defined the ratio of driving force versus velocity [12]. In this regard, the definition of characteristic impedance in clinical studies and in the tube-load model is not correct, in the sense that pulsatile pressure is not driving force, and blood flow rate is not velocity. The arterial tree contains multiple arterial segments with a relatively sudden radius change at segment connects [1,2,9]. Furthermore, an arterial segment contains a tapered geometry (slowly reduced radius from the entrance to the exit) [1,2,7,9]. Here, the effect of the arterial geometrical change on wave reflection is examined via characteristic impedance and load impedance.
As far as there is a reduction in arterial radius at a location, there exists impedance mismatch and then wave reflection occurs [12]. This equation explains wave reflection taking place everywhere along the length of a tapered arterial segment, which is observed in the tapered tube-load and the distributed 1D models [1]. Thus, the forward and reflected waves measured at an artery site need to be treated as the compound forward and reflected waves, which result from wave reflection at multiple sites in the arterial tree [1].
Load impedance at periphery ZL results from 1) impedance mismatch between the aorta-periphery section and rest arteries and 2) boundary conditions at termination [13]. Neither a pure resistor nor a three-element Windkessel model is capable of capturing these two factors. In the context of the 1D wave propagation, Re(ZL) presents the part of energy transmitted to rest arteries and Im(ZL) represents the part of energy reflected back to the aorta-periphery section [12]. A pure resistor as ZL indicates no wave reflection at periphery, which is physiologically unrealistic. Furthermore, this pure resistor gives rise to harmonic-independence of ZL. Combination of Eqs. (44a) and (48) explains the oscillation of input impedance with harmonics, which is observed in the tube-load model with a pure resistor as the load [1]. The Windkessel model as ZL allows wave reflection and harmonics-dependence. However, the dependence of ZL on harmonics is stipulated by the values of the three elements, as shown in Eq. (44b). It is well known in the acoustical field that a tapered geometry causes the dependence of boundary conditions on harmonics [12]. Owing to mathematical complexity, an explicit expression on harmonics-dependence of boundary conditions for a tapered geometry is unavailable. Anatomical complexity in the arterial tree exacerbates difficulty in attaining harmonic-dependence of boundary conditions at termination. However, As shown in Fig. 4, harmonics-dependence of ZL can be obtained from the measured Z0 in clinical studies.
5.3 Measured Input Impedance, Return Time, Reflection Magnitude, and Augmentation Index Versus Harmonics.
Based on Fig. 4, input impedance is determined by load impedance and arterial properties and geometries in the aorta-periphery section. Although harmonics-dependence of wave velocity and characteristic impedance in the aorta-periphery section is clear, harmonics-dependence of load impedance is unknown, as explained in Sec. 5.2. Thus, how the input impedance varies with harmonics can only be obtained from the measured values. As an example, the measured harmonics of pulsatile pressure and blood flow waveform at the aorta in the literature [8], as shown in Figs. 5(a) and 5(b), are used to calculate the measured input impedance based on Eq. (31). The variation of the measured input impedance with harmonics is plotted in Fig. 5(c). Harmonics-dependence of load impedance can then be estimated from Eq. (23b).
It is found in clinical studies that the measured return time of the reflected waveform ηab(t) is longer than 2 L/c0 [1]. As shown in Eq. (26b), the complex value of ZL leads to a phase delay and thus extra time delay. Since it is found that return time depends little on c0 in clinical studies [1], it might indicate that high arterial stiffness leads to a higher θn at periphery. Because ZL and round-trip wave transmission are embedded in Z0, input impedance Z0 is allowed to estimate reflection magnitude and return time. As shown in Fig. 6(a), reflection magnitude varies with harmonics, and reflection magnitude for the whole waveform is not the average of the ten harmonics. As expressed in Eq. (28), reflection magnitude of the whole waveform is determined by not only ZL (or Z0) but also harmonics of the forward waveform. As shown in Fig. 6(b), due to harmonics-dependence of phase delay at periphery and wave velocity, the reflected waves of different harmonics arrive at the aorta at different return times. As shown in Fig. 7, harmonics-dependence of return time makes it elusive to identify the foot of the reflected waveform [1,14]. Defined as the ratio of augmented pressure to pulsatile pressure, AP/PP [1,14], AI varies greatly with the location of this foot. As a composite indicator of return time and reflection magnitude, AI also manifests harmonics of aortic pressure waveform. Since return time, reflection magnitude, and AI are all affected by harmonics of aortic pressure waveform, the reconstructed aortic pressure waveform needs to capture accurate harmonics information for accurate estimates of these clinical indices. Given the dependence of return time and reflection magnitude on arterial stiffness (i.e., c0), arterial radius, as well as harmonics, it is no surprise that AI carries independent clinical values, as compared with arterial stiffness [2,5]. Taken together, the definitions of return time for the whole waveform and AI are inconsistent with harmonics-dependence of return time, and reflection magnitude for the whole waveform and AI carry harmonics information of aortic pressure waveform. Given the dominance of the 1st harmonic in aortic pressure waveform, reflection magnitude and return time of the 1st harmonic might serve better as clinical indices for arterial properties and geometries and wave reflection at periphery, and harmonics of the whole waveform might serve as a clinical index indicative of the LV function, instead of AI.
5.4 The Left Ventricle-Artery Interaction for Driving Force on the Left Ventricle and Harmonics of Aortic Pressure Waveform.
To calculate the driving force on the LV based on the lumped-element model, input impedance defined in the vibrating-string model is calculated from the measured input impedance, as shown in Fig. 8. Based on the harmonics of aortic pressure waveform in Fig. 5(a), the two driving forces on the LV is calculated from Eq. (36a) and Eq. (36b) and are plotted in Figs. 9(a) and 9(b). The value of M for the LV is chosen to be 0.3 kg (close to the mass of the heart), and the value of K is chosen such that the LV operates at resonance at the heartbeat. It is assumed that the LV operates with high efficiency so that the value of D is very small. Figure 10(a) compares the phases of blood velocity, radial wall displacement (or pulsatile pressure), and Fu(t). It is interesting to note that Fu(t) is similar to the waveform of an ECG signal to some extent [15] and is also ahead of blood velocity, similar to an ECG signal. Figure 9(c) shows the calculated driving force based on Eq. (50), which is completely different from the waveform of an ECG signal. As shown in Fig. 10(b), this driving force is in phase with blood velocity. As such, the tube-load model might not be suitable for studying the LV-artery interaction for relating aortic pressure waveform to the LV function.
As shown in Fig. 3, the LV and the arterial tree and its termination all play a role in determining harmonics of ηa(t). It is well established in the musical instruments field [16] that the excitation source (e.g., a soft hammer versus a rigid hammer), Zc in the vibrating string, and ZL at the other end of the string will affect the interaction between the excitation source and the vibrating string, and consequently the generated harmonic amplitude percentages of ηa(t) at different n-levels. With the excitation source at one end, the generated harmonics of ηa(t) usually contain a decreasing trend of amplitude percentages with higher n-levels [16], as shown in Fig. 5(a). Given the role of the LV in determining harmonics of aortic pressure waveform and the role of harmonics in AI, it is no surprise that AI is found to manifest not only arterial properties and geometries but also the LV function [17].
The model for the LV-artery interaction in Fig. 3 is a rather simplified model to relate the LV to the arterial tree. Similar to hammer-string interaction in a piano [16], it is exceedingly intricate to capture the LV-artery interaction for predicting harmonics of aortic pressure waveform. Generally speaking, low Zc increases the amplitudes of higher harmonics, while high Zc increases the amplitudes of lower harmonics [18]. Higher harmonics lead to higher amplitudes of wall shear stress, as compared with lower harmonics [10]. This may explain the importance of accurate representation of higher harmonics in reconstructed aortic pressure waveform, which dictates the peak wall stress in the LV, with the latter being a fundamental biophysical driver of myocardial hypertrophy [5].
5.5 Machine-Learning Techniques for Reconstruction of Aortic Pressure Waveform.
In the studies of the tube-load model, machine learning techniques have focused on optimizing the values of load impedance to improve accuracy in reconstructed aortic pressure waveform from its peripheral counterpart, to the neglect of physical implications of load impedance and without examining the input impedance [2–7]. This explains why a pure resistor is used as the load [4] and a generic pole–zero model for the load [3] is considered to carry no physiological meaning. Since a pure resistor is inconsistent with wave reflection, it is impossible to extract physiological meaning from the obtained results based on it [1,4]. This work shows that the essence of these machining-learning techniques is about chasing physiologically realistic harmonic-dependence of load impedance [3,5]. Yet, given that at least 20 unknown values (if only the first ten harmonics are utilized.) are needed for accurate representation of harmonic-dependence of load impedance, there are a great many combinations of the 20 values to match the reconstructed one with the measured one. Taken together, the results derived from application of machine-learning techniques to reconstruction of aortic pressure waveform do not carry much physiological meaning unless the application is consistent with the associated theories and physical implications.
5.6 Study Limitations.
In this work, the aorta-periphery section is treated as a uniform vibrating string with the same arterial geometries and properties. This treatment neglects the following three factors: 1) tapered arterial geometry and nonuniform arterial properties in an arterial segment, 2) different geometries and properties among multiple arterial segments, and 3) bifurcations. These neglected factors are identical to those for the uniform tube-load model and those for the calculation of PWV in clinical studies [1–3]. However, as compared with the tube-load model and PWV in clinical studies, harmonics-dependent wall shear stress is included in this work, given that it causes non-negligible dependence of wave transmission and reflection on harmonics and thus greatly affects aortic pressure waveform.
Since the measured input impedance captures the effect of the neglected factors on wave reflection and load impedance, the effect of the neglected factors on wave reflection is factored in load impedance in Eq. (52). In clinical studies, pulsatile pressure waveform is treated as one forward waveform and one reflected waveform for extracting clinical measures, without considering repeated wave reflections at multiple sites. The uniform vibrating-string model also neglects repeated wave reflection and provides one forward waveform and one reflected waveform. Finally, arterial properties and geometries in the vibrating-string model need to take equivalent values for accounting for their variations among multiple arterial segments.
6 Conclusion
In light of reconstruction of aortic pressure waveform with accuracy for its clinical indices, a vibrating-string model is developed to examine the roles of input impedance, load impedance, and harmonics in aortic pressure waveform and its three clinical indices. On the one hand, the revealed role of the three parameters in the clinical indices identifies the importance of physiologically realistic load impedance in reconstruction of aortic pressure waveform with accuracy. On the other hand, this work reveals physical implications and physiological implications of the three clinical indices and identifies a lack of consideration of harmonics-dependence of wave transmission and reflection in their definitions. Harmonics of aortic pressure waveform is suggested as a clinical index indicative of the LV function, instead of AI.
This work provides a theoretical guidance for reconstruction of aortic pressure waveform and interpretation of aortic pressure waveform for assessing arterial health and the LV function. The key insights shed by this work on reconstruction of aortic pressure waveform and its associated physical implications are summarized here:
Load impedance must be complex, with the real part representing energy transmitted and the imaginary part representing energy reflected back. Complex load impedance translates to non-zero phase angle of reflection coefficient, which represents the phase delay of the reflected wave relative to the forward wave at periphery and affects return time at the aorta.
A three-element Windkessel model in the tube-load model stipulates harmonic-dependence of load impedance, which is inconsistent with physiological realities.
Input impedance is a dependent parameter, and is determined by load impedance and arterial geometries and properties. Physiologically realistic harmonic-dependence of load impedance can only be obtained from the measured input impedance, instead of machine-learning techniques. Return time at the aorta can be estimated from input impedance, due to the dependence of input impedance on load impedance.
To capture the LV-artery interaction for aortic pressure waveform, the arterial tree and its termination need to be treated with the real part of input impedance as a damper and the imaginary part of input impedance as a spring. Together with arterial properties and geometries, the LV plays a role in determining harmonics of aortic pressure waveform.
Fluid loading causes harmonic-dependence of wave velocity and characteristic impedance. Accordingly, each harmonic has its own return time and reflection magnitude at the aorta. Meanwhile, harmonics of the forward pressure waveform at the aorta are non-uniform. As such, harmonics of aortic pressure waveform play an important role in determining the three clinical measures: return time, reflection magnitude, and AI. AI carries much more physiological information than arterial stiffness, and is indicative of the LV function.
Based on this work, it might be more practical to develop a better understanding of harmonic-dependence of input impedance and harmonics of the forward pressure waveform at the aorta for different groups of population, based on the measured blood flow rate and the measured aortic pressure waveform, prior to reconstruction of aortic pressure waveform from peripheral counterpart for estimates of its clinical measures using a 1D model.
Funding Data
Directorate for Engineering (Funder ID: 10.13039/100000084).
Data Availability Statement
The authors attest that all data for this study are included in the paper.