Orthopaedic research on in vitro forces applied to bones, tendons, and ligaments during joint loading has been difficult to perform because of limitations with existing robotic simulators in applying full-physiological loading to the joint under investigation in real time. The objectives of the current work are as follows: (1) describe the design of a musculoskeletal simulator developed to support in vitro testing of cadaveric joint systems, (2) provide component and system-level validation results, and (3) demonstrate the simulator’s usefulness for specific applications of the foot-ankle complex and knee. The musculoskeletal simulator allows researchers to simulate a variety of loading conditions on cadaver joints via motorized actuators that simulate muscle forces while simultaneously contacting the joint with an external load applied by a specialized robot. Multiple foot and knee studies have been completed at the Cleveland Clinic to demonstrate the simulator’s capabilities. Using a variety of general-use components, experiments can be designed to test other musculoskeletal joints as well (e.g., hip, shoulder, facet joints of the spine). The accuracy of the tendon actuators to generate a target force profile during simulated walking was found to be highly variable and dependent on stance position. Repeatability (the ability of the system to generate the same tendon forces when the same experimental conditions are repeated) results showed that repeat forces were within the measurement accuracy of the system. It was determined that synchronization system accuracy was and was based on timing measurements from the robot and tendon actuators. The positioning error of the robot ranged from to , depending on measurement condition (e.g., loaded or unloaded, quasistatic or dynamic motion, centralized movements or extremes of travel, maximum value, or root-mean-square, and -, - or -axis motion). Algorithms and methods for controlling specimen interactions with the robot (with and without muscle forces) to duplicate physiological loading of the joints through iterative pseudo-fuzzy logic and real-time hybrid control are described. Results from the tests of the musculoskeletal simulator have demonstrated that the speed and accuracy of the components, the synchronization timing, the force and position control methods, and the system software can adequately replicate the biomechanics of human motion required to conduct meaningful cadaveric joint investigations.
Skip Nav Destination
e-mail: davisb3@ccf.org
Article navigation
February 2010
Design Innovations
Design and Validation of a General Purpose Robotic Testing System for Musculoskeletal Applications
Lawrence D. Noble, Jr.,
Lawrence D. Noble, Jr.
Department of Biomedical Engineering, Lerner Research Institute, and Orthopaedic and Rheumatologic Research Center,
Cleveland Clinic
, Cleveland, OH 44195
Search for other works by this author on:
Robb W. Colbrunn,
Robb W. Colbrunn
Department of Biomedical Engineering, Lerner Research Institute, and Orthopaedic and Rheumatologic Research Center,
Cleveland Clinic
, Cleveland, OH 44195
Search for other works by this author on:
Dong-Gil Lee,
Dong-Gil Lee
Department of Biomedical Engineering, Lerner Research Institute, and Orthopaedic and Rheumatologic Research Center,
Cleveland Clinic
, Cleveland, OH 44195; Department of Orthopaedics and Sports Medicine, University of Washington
, Seattle, WA 98195
Search for other works by this author on:
Antonie J. van den Bogert,
Antonie J. van den Bogert
Department of Biomedical Engineering, Lerner Research Institute, and Orthopaedic and Rheumatologic Research Center,
Cleveland Clinic
, Cleveland, OH 44195
Search for other works by this author on:
Brian L. Davis
Brian L. Davis
Department of Biomedical Engineering, Lerner Research Institute, and Orthopaedic and Rheumatologic Research Center,
e-mail: davisb3@ccf.org
Cleveland Clinic
, Cleveland, OH 44195
Search for other works by this author on:
Lawrence D. Noble, Jr.
Department of Biomedical Engineering, Lerner Research Institute, and Orthopaedic and Rheumatologic Research Center,
Cleveland Clinic
, Cleveland, OH 44195
Robb W. Colbrunn
Department of Biomedical Engineering, Lerner Research Institute, and Orthopaedic and Rheumatologic Research Center,
Cleveland Clinic
, Cleveland, OH 44195
Dong-Gil Lee
Department of Biomedical Engineering, Lerner Research Institute, and Orthopaedic and Rheumatologic Research Center,
Cleveland Clinic
, Cleveland, OH 44195; Department of Orthopaedics and Sports Medicine, University of Washington
, Seattle, WA 98195
Antonie J. van den Bogert
Department of Biomedical Engineering, Lerner Research Institute, and Orthopaedic and Rheumatologic Research Center,
Cleveland Clinic
, Cleveland, OH 44195
Brian L. Davis
Department of Biomedical Engineering, Lerner Research Institute, and Orthopaedic and Rheumatologic Research Center,
Cleveland Clinic
, Cleveland, OH 44195e-mail: davisb3@ccf.org
J Biomech Eng. Feb 2010, 132(2): 025001 (12 pages)
Published Online: January 28, 2010
Article history
Received:
December 18, 2006
Revised:
October 15, 2009
Posted:
December 17, 2009
Published:
January 28, 2010
Online:
January 28, 2010
Citation
Noble, L. D., Jr., Colbrunn, R. W., Lee, D., van den Bogert, A. J., and Davis, B. L. (January 28, 2010). "Design and Validation of a General Purpose Robotic Testing System for Musculoskeletal Applications." ASME. J Biomech Eng. February 2010; 132(2): 025001. https://doi.org/10.1115/1.4000851
Download citation file:
Get Email Alerts
Simulating the Growth of TATA-Box Binding Protein-Associated Factor 15 Inclusions in Neuron Soma
J Biomech Eng (December 2024)
Effect of Structure and Wearing Modes on the Protective Performance of Industrial Safety Helmet
J Biomech Eng (December 2024)
Sex-Based Differences and Asymmetry in Hip Kinematics During Unilateral Extension From Deep Hip Flexion
J Biomech Eng (December 2024)
Related Articles
Ground Reaction Forces and Lower Extremity Kinematics When Running With Suppressed Arm Swing
J Biomech Eng (December,2009)
Musculoskeletal Modeling and Dynamic Simulation of the Thoroughbred Equine Forelimb During Stance Phase of the Gallop
J Biomech Eng (April,2005)
Use of Virtual, Interactive, Musculoskeletal System (VIMS) in Modeling and Analysis of Shoulder Throwing Activity
J Biomech Eng (June,2005)
Design of an Endoreactor for the Cultivation of a Joint-Like-Structure
J. Med. Devices (June,2009)
Related Proceedings Papers
Related Chapters
An Adaptive Fuzzy Control for a Multi-Degree-of-Freedom System
Intelligent Engineering Systems Through Artificial Neural Networks, Volume 17
Servo System Position Control of Honing Machine Based on Fuzzy Control
International Conference on Mechanical and Electrical Technology, 3rd, (ICMET-China 2011), Volumes 1–3
Intuitive Optimization
Engineering Optimization: Applications, Methods, and Analysis